Development of a cardiovascular simulator incorporating a phantom simulation of the arterial system : [a thesis submitted to Auckland University of Technology in partial fulfilment of the requirements for the degree of Master of Engineering (ME), 2021] / Liam Foley ; supervisors: Andrew Lowe, Martin Stommel.

A mock circulatory loop was constructed to test a new blood pressure device under development. The loop was intended to replicate a range of cardiac conditions with the implementation of an artificial arterial tree. An in-depth literature review was conducted on the previous construction on mock cir...

Full description

Saved in:
Bibliographic Details
Main Author: Foley, Liam (Author)
Corporate Author: Auckland University of Technology. School of Engineering, Computer and Mathematical Sciences
Format: Ethesis
Language:English
Subjects:
Online Access:Click here to access this resource online
Description
Summary:A mock circulatory loop was constructed to test a new blood pressure device under development. The loop was intended to replicate a range of cardiac conditions with the implementation of an artificial arterial tree. An in-depth literature review was conducted on the previous construction on mock circulatory loops and arterial phantoms. The methods employed previously were analysed and implemented in this project. Lost core casting methods were used to create complex artery phantoms, using 3D printed moulds and silicone elastomers, yielding an arterial tree that was dimensionally and functionally accurate in reproducing pulse wave amplification and arterial wall compliance. A synthetic ventricle pump was developed in conjunction with a variable compliance chamber and peripheral resistance. The pump consists of a synthetic ventricle inside a water-filled chamber with a motor actuating a piston to change the volume, successfully reproducing the flow desired. The compliance chamber emulated a bladder accumulator to reproduce capacitances as desired, and a motor actuated gate valve simulates peripheral resistance. The overall system successfully reproduced human pressures and flows, with sufficient damping being observed due to the compliance chamber and the elastic vessels.
Author supplied keywords: Cardiovascular; Mock Circulatory Loop; Phantom Arteries; Blood Pressure.
Physical Description:1 online resource
Bibliography:Includes bibliographical references.
Requests
Request this item Request this AUT item so you can pick it up when you're at the library.
Interlibrary Loan With Interlibrary Loan you can request the item from another library. It's a free service.